Olympus BX53-P is a polarizing microscope for working with substances that transmit light differently in different directions. Minerals in the shelf, crystals, fibers, polymer films in polarized light change color and brightness when rotated, and these changes can be identified and measured.

The model is based on the BX3 platform with infinitely adjusted UIS2 optics and is designed for petrography, mineralogy, chemistry and materials science. Next, we'll talk about what a polarizing microscope shows, how orthoscopy differs from conoscopy, why compensators are needed, and how to prepare a sample.

What is visible in polarized light

In a polarizing microscope, light passes through two filters. The polarizer under the sample transmits vibrations in only one plane. The analyzer above the sample is turned at a right angle to it. If there is nothing between them or the substance is isotropic, like glass, the field remains dark.

An anisotropic substance, such as quartz or calcite, splits light into two beams at different speeds. After the analyzer, they interfere, and the crystal becomes light or colored. The observer receives several signs to determine:

  • Interference coloring. It depends on the thickness of the sample and the strength of the birefringence of the substance.
  • Fading out. When the table is rotated, the crystal turns dark four times per revolution. The angle of extinction relative to the faces helps to distinguish minerals.
  • Pleochroism. Color change during rotation in one polarizer, typical, for example, for biotite and hornblende.
  • Twinning and zoning. Visible as stripes and zones with different colors inside the grain.

Therefore, a polarizing microscope requires a rotating table with an angular scale. It is used to calculate the extinction angles and the orientation of the grains.

Orthoscopy and conoscopy

In the normal orthoscopic mode, the microscope shows an image of the sample itself: grains, their shape, and color. This is sufficient for most descriptions of the slot.

Conoscopy is used to determine the optical properties of the crystal. A Bertrand lens is inserted into the course of the rays, and instead of the grain image, the observer sees an interference pattern: crosses and rings. It determines whether a crystal is uniaxial or biaxial, and the optical sign of the mineral. Conoscopy requires a high-aperture condenser and a high-magnification lens.

For the BX53-P, the U-CPA intermediate nozzle switches between modes. With it, they switch from orthoscopy to conoscopy without rebuilding the microscope.

Compensators and stroke difference measurement

Interference coloring allows you to estimate the difference in the course of two rays per eye on a color scale. Compensators are used for accurate values. These are plates with a known stroke difference.,